E201-C or E201-F (water proof BNC connector)is most suitable. Both are refillable. Reply from TB sellers, 雷磁 brand E201-C response time is typically 1x to 2x seconds. So the main difference is just the good looking white casing.
This pH combination electrode is made of sensitive glass membrane with low impedance.It can be used in a variety of PH measurements with fast response, good thermal stability.It has the good reproducibility, difficult to hydrolysis, and basically eliminate the alkali error .In 0 to 14pH range,the output voltage of the electrode is linear.The reference system which consist of the Ag/AgCl gel electrolyte salt bridge has a stable half-cell potential and excellent anti-pollution performance.The ring PTFE membrane is not easy to be clogged, so the electrode is suitable for long-term online detection.
Googling up from internet, it reveals:
Haoshi 東莞豪斯H-101 , used in DFRobot pH meter pro v2.¥140. detailed spec missing. Armor casing. PTFE ring junction. Some more info here
1. Low impedance coating or fast response is more like marketing. Such chemical reactions is slow anyways. Take 15 seconds or more is common.
2. PTFE ring is quite common material use. Nothing special.
3. Long term operation. is another marketing. As electrolytes ( KCL)is constantly ( slowly) leaking out, it will be used up eventually. If it is non refillable type. When you use up the electrolytes, You have no choice but to buy a new one. The leak rate is dependent on reference junction type
4. It has 25mm O.D. NPT/ PG threads , long 5m cable and ABS casing, make it easy to install and extend the length. Eg for commercial environment.
5. Need to be understand clearly your application need , to say if it's suitable for industrial use ( see above)
6. Most important in industrial applications, is to have electrical isolation. pH , EC probe can interfere pH measurement. Siemens forum , which is not designed/included in this DFRobot industrial grade kit. Radicaldiy
Question: adding 1/4 teaspoon of 10% phosphoric acid to 19 litres of water, what will be the pH? Let's go through the calculations step by step. Step 1. Convert mass ( or volume) of 10% phosphoric acid to mole per litre ( mol/L) The above calculation shows that 1/4 tsp of H3PO4 acid is 6.97e-5 mol/L Step 2. Input this value to BATE (or other app) to calculate pH The answer: pH = 4.16, ie too acidic !? If you use pipette instead of teaspoon, 1 drop = 0.05ml. So 1/4 TSP means roughly 25 drops If you actually aim for pH 5.6 (which is optimum pH for mashing), conc will be 2.51 e-6, or 1/100 tsp , 0.044 ml, roughly 1 drop Let's roll back a bit. From recipes of BarleyWine , summer rye , hefe Gordon Strong suggested to use 1/4 teaspoon of 10% phosphoric acid, which is targeting pH of 5.5..... Discussion on homebrewtalk forum. However using online BF water calc , Sparge pH would be 4.6. but he also added 1 tsp of CaCl2. 2H20. Mash pH change from 5.7 t...